Troubleshooting Flow Cytometry
Troubleshooting
This troubleshooting page gives the problem, possible cause and suggested solution for problems during the flow cytometry application:
Problem: No signal or weak fluorescence intensity
Incorrect storage | Ensure that all antibodies have been stored correctly according to the manufacturer’s instructions. |
The antibodies are too old | Make sure that the antibodies have not exceeded their date of expiration. |
The incubation time and temperature are not optimal | Optimize the incubation time and temperature of the primary and secondary antibody. |
The signal is not correctly compensated | Check whether the positive single colour control is set up correctly on the flow cytometer and is compensated correctly to capture all the events. |
There is insufficient antibody present for detection | Antibodies are over diluted, make sure to use the correct concentration by titrating the antibodies before use. |
The intracellular target is not accessible | Check if the target protein is really intracellular. Ensure adequate permeabilization for internal staining. To stop all reactions to prevent internalisation of cell surface proteins, all the steps should be done at 4°C, with ice cold reagents. |
The fluorochrome conjugate is too large for intracellular staining | The fluorochromes used for intracellular staining experiments should have low molecular weight. Larger flourochromes can reduce the antibody motility and possibly its entry into the cell. |
The lasers are not aligned | Make sure that the lasers on the flow cytometer are aligned correctly by running flow check beads and adjust the alignment if necessary. You may need to consider having the machine serviced when the lasers do not align correctly or if drift occurs. |
The target protein is not present or the expression is too low | Check the literature for antigen expression and incorporate a positive control of the known antigen expression alongside test material. If antigen expression is weak, select an antibody that is conjugated to a brighter fluorochrome. |
The target protein might be soluble/secreted. | To be detected easily by flow cytometry, the target protein needs to be membrane bound or cytoplasmic. For intracellular staining a golgi-block step, such as with Brefaldin A, may improve the signal achieved |
The antigen is not recognized by the antibody | Check if the antibody will react with the species being used. Not all antibodies will cross-react across species. |
The offset is too high or the gain is too low | A positive control should be used to set up the flow cytometer correctly. Use the offset to make sure that the fluorescent signal from cells is not being cut off. Increase the gain to increase the signal but do this within reason |
The fluorescence of the fluorochrome has faded | The antibody may have been kept in the light too long, a fresh antibody is required. Use a new batch of antibodies. |
The primary and the secondary antibody are not compatible | Use a secondary antibody that was raised against the species in which the primary was raised (e.g primary is raised in rabbit, use anti-rabbit for secondary antibody). |
Secondary antibody is not active/not conjugated to a fluorochrome | Make sure that the secondary antibody is active. Ensure that the appropriate fluorochrome-conjugated secondary antibody is used. |
Problem: The fluorescence intensity is too high
The antibody concentration is too high | A high concentration of antibody will give high non-specific binding or very high intensity of fluorescence. The amount of antibody added to each sample should be reduced. |
The excess antibody is trapped | This can be a particular problem in the intracellular staining where large fluorochrome molecules on the antibody can be trapped. Ensure that adequate washing steps are taken and include Tween or Triton in the wash buffers. |
Problem: There is a high background
The gain set is too high, the offset is too low | Use the positive control to set up the flow cytometer correctly again, using the offset to reduce background from small particles and reduce the gain to decrease the signal but do this within reason. |
Excess antibody | The antibody concentration should be decreased. You can also add a detergent to the wash buffers to ensure washing away of excess antibody. |
Problem: Non-specific staining
The sample gives autofluorescence | The level of autofluorescence can be checked by including a tube of cells without any antibody into your panel. |
Non-specific staining may be due to the secondary antibody | Select a secondary antibody that will not cross-react with the target sample. |
The washing is incorrect | Make sure that correct washing steps are included. |
Problem: Two or more cell populations are observed when there should be only one population
There is more than one cell population present which is expressing the target protein | Check the expected expression levels from the cell types that are present in the sample and ensure adequate cell separation if necessary. |
Cell doublets are present in the sample | Cell doublets will show as a second cell population at approximately twice the fluorescence intensity on the plot. Mix the cells gently before the staining process and again before running them on the cytometer using a pipette. Sieve or filter cells to remove clumps (30 µl Nylon Mesh) |
Problem: High side scatter background (from small particles)
The cells are lysed | A high side scatter background can be caused by cells in the sample that have been lysed and broken up. Samples should be fresh and prepared correctly, so do not centrifuge at a high rotor speed or vortex too violently. |
Bacterial contamination | Make sure that the sample is not contaminated. Bacteria will auto fluoresce at a low level, and will give a high event rate. |
Unusual scatter profile | The cells must be used as fresh as possible. The scatter profile may be showing dead cells and debris. Activation methods may affect scatter characteristics of cells. |
Problem: Low event rate
Number of cells/ml too low | Run the sample with 1 x 106 cells/ml. Make sure that the cells are mixed well, but gently. |
The cells are clumped and blocking the tube | Pipette the sample gently several times before staining for a homologous single cell suspension. Make sure you mix the sample again before the running process. In extreme cases, sieve or filter the cells to remove clumps (30 µl Nylon Mesh). |
Problem: High event rate
Number of cells/ml too high | Dilute the sample to 1×105 or 1×106 cells/ml. |
We are glad to support you!
Take advantage of our dedicated support team for any technical assistance you need while using our products or considering them for your research needs.